Pressure measurement |
Operating pressure ranges |
PDCR 1800 (mV)
0.75, 1.5 mH2O gauge, 3.5, 7, 10, 15, 20, 35, 50, 70, 100, 150, 200, 350, 600 mH2O gauge and absolute
Other units may be specified |
Overpressure |
The operating FS pressure range may be exceeded by the following multiples with negligible effect on calibration:
- 8 x for ranges up to 1.5 mH2O
- 6 x for ranges above 1.5 to 3.5 mH2O
- 4 x for ranges above 3.5 mH2O (1400 mH2O max.)
|
Pressure containment |
- 10 x for ranges up to 3.5 mH2O gauge
- 6 x for ranges above 3.5 mH2O gauge (1400 mH2O
maximum)
- 200 bar for absolute ranges
|
Media compatibility |
Fluids compatible with titanium (body), acetyl (nose cone) and polyurethane or Hytrel® 6108 (cable assembly) |
Excitation voltage |
PDCR 1800 (mV)
10 V at 5 mA nominal
Output is ratiometric to supply within 2.5 V to 12 V limits. |
Pulse power excitation |
Recommended power-on time before output sample:
10 ms
|
Output signal |
PDCR 1800
- 25 mV for 0.75 mH2O range
- 50 mV for 1.5 and 3.5 mH2O ranges
- 100 mV for ranges 7 mH2O and above
|
Performance specification |
Accuracy |
Combined effects of non-linearity, hysteresis and repeatability:
- Standard: ±0.1% FS BSL maximum
- Option D: ±0.06% FS BSL maximum (±0.08% FS BSL max. for 1 mH2O and below)
|
Zero offset and span setting |
PDCR 1800
- Typical: ±1.5 mV
- Maximum: ±3 mV
|
Long-term stability |
±0.1% FS per annum |
Operating temperature range |
-20 to 60°C (-4 to 140°F) |
Compensated temperature range |
-2 to 30°C (28 to 86°F) |
Temperature effects |
±0.3% FS Temperature Error Band (TEB) for 3.5 mH2O range and above
±0.6% FS TEB for ranges below 3.5 mH2O |